#5602
Surface area analyzer
QUANTACHROME – NOVA 1200E
DESCRIPTION
Automatic surface area and pore size analyzer based on gas adsorption, with a single analysis station. The 1200e version accepts adsorptive gases other than nitrogen, which enables micropore characterization.
MAIN FEATURES
- Analysis stations: 1; built-in degassing ports: 2
- “Any Gas” capability: accepts N2, Ar, CO2, H2, light hydrocarbons and other non-corrosive gases
- Surface area range: from 0.01 m²/g, with no known upper limit
- Pore size range: 0.35 to more than 400 nm
- Minimum detectable pore volume: 2.2 × 10^(-6) mL/g (liquid)
- Degassing: by vacuum or flow, from ambient temperature to 450 °C in 1 °C steps (350 °C with standard mantles)
- Pressure transducer: accuracy ±0.1 % of span; resolution 2 × 10^(-5) in P/P0
- NOVA technology: helium-free analysis (NO Void Analysis), or classical helium void volume mode
- Data: up to 100 adsorption and 100 desorption points; NOVAWin software with BET, Langmuir, t-plot, BJH, DR/DA, MP and DFT
MAIN APPLICATIONS
- BET specific surface area determination of catalysts and supports
- Characterization of activated carbons and of their micropore development
- Evaluation of adsorbents for water treatment and gas separation
- Pore size distribution of silicas, clays and mesoporous materials
ASSOCIATED TECHNOLOGY AND SERVICES